Orifice Plate Flow & Differential Pressure Calculator
Compare two concentric orifice bores for steady single-phase liquid in a full circular pipe. Use your calibrated discharge coefficients, their bounds and their valid differential-pressure intervals. The worksheet shows where both the calibration and gauge cover the full range, and projects the differential pressure a flow target would require.
Compare calibrated orifice plates
Check whether the complete reading or projected target fits your calibration and gauge.
Edit meter and calibration
Use the Cd and pressure range calibrated with each installed plate and its pressure taps. Enter absolute half-widths for reading and Cd.
Plate A
Nominal flow
146.23 L/min
Full coveragePlate B
Nominal flow
256.56 L/min
Full coverageReading 12 ± 0.25 kPa. Full flow bounds require the entire pressure interval to fit the calibration and gauge.
Flow and pressure envelope
Curves and coefficient bands stop at the usable pressure limits. The horizontal axis ends at the gauge span.
| Range check | Plate A | Plate B |
|---|---|---|
| Flow interval | 142.33 to 150.17 L/min | 247.58 to 265.65 L/min |
| Calibrated range | 2 to 28 kPa | 1 to 19 kPa |
| Usable range | 2 to 25 kPa | 1 to 19 kPa |
| Calibration coverage | Full coverage | Full coverage |
| Gauge coverage | Full coverage | Full coverage |
| Diameter ratio (β) | 0.4 | 0.525 |
Select audit CSV
For steady single-phase liquid in a full pipe. Use the pressure difference at the calibration taps. Operating and installation limits come from the meter documentation and calibration.
How to use this calculator
- Enter the inside diameter of the full pipe, liquid density, differential-pressure reading and its absolute error bound. Enter the upper end of a zero-based gauge span.
- For each plate, enter its bore, the calibrated dimensionless Cd, an absolute coefficient bound and the differential-pressure interval where that calibration applies.
- Use Measured flow to see the nominal estimate and whether the complete reading interval fits both the calibration and the gauge.
- Enter a target in L/min and open Target pressure. Compare the projected pressure interval with each plate's usable span. Download both cases and the source inputs as CSV.
The Formula
The mathematical logic powering this calculation is:
Let D be the pipe inside diameter, d the concentric bore, β = d/D, A = πd²/4 in m² and ρ the entered liquid density in kg/m³. For steady single-phase liquid in a full circular pipe, the meter model is Q = Cd·A·√[2ΔP / (ρ(1−β⁴))], with ΔP in Pa measured between the calibration's taps. Q is in m³/s. The worksheet converts kPa to Pa and m³/s to L/min. It sets the expansion factor to 1 for its incompressible model. The tap differential is distinct from permanent pressure loss.
Cd is a dimensionless discharge coefficient on this exact basis, retaining the velocity-of-approach factor 1/√(1−β⁴). Enter a coefficient and bounds calibrated for the actual bore, pipe, taps, fluid and installed flow conditions. A coefficient that already absorbs the approach factor, a valve Cv, or a gas expansion factor uses a different basis. The interval applies only over the entered calibrated differential pressures. Pipe condition, Reynolds number and flow profile can change the coefficient.
For a ΔP reading p with absolute bound e, possible pressures are [p−e,p+e]. The usable pressure range is the intersection of the entered calibration interval and the zero-to-span gauge range. The nominal estimate is available when p is usable. A complete flow interval is available only if the whole possible pressure interval is usable. The lower/upper flow estimates pair low/low and high/high pressure and Cd. The model does not clip an out-of-range reading or assign it a flow.
Write Q = k√ΔP where k includes the chosen Cd and constants. A target Q requires ΔP = (Q/k)². At the two Cd bounds, the higher coefficient gives the lower required pressure. Values outside calibration are conditional projections only. Coverage indicates whether all, some or none of the projected pressure interval overlaps the usable range. The checked result is a bounded sensitivity calculation for entered independent endpoints and fixed density and geometry. It is not a confidence interval and excludes uncertainties in diameter, density or installation.
Enter positive pipe and bore diameters through 5,000 mm with d < D, density through 50,000 kg/m³, Cd bounds above zero through 2, and pressure fields through 1,000,000 kPa with gauge span and calibration maxima positive. Target flow is 0 through 10,000,000 L/min. Fields accept up to six decimal places. These are numerical input limits. The range and assumptions apply to steady liquid that remains single phase. Gas, steam, flashing, cavitation, pulsation, installation checks and ISO 5167 compliance require separate evaluation.Worked Example
If you input:
- Pipe and liquid: 80 mm inside diameter, 998 kg/m³ density
- Measured ΔP: 12 ± 0.25 kPa
- Plate A: 32 mm bore, Cd 0.61 ± 0.01 calibrated at 2–28 kPa
- Plate B: 42 mm bore, Cd 0.605 ± 0.015 calibrated at 1–19 kPa
- Target and gauge: 220 L/min target, 25 kPa gauge span
Result:
The nominal readings are 146.23 L/min for Plate A and 256.56 L/min for Plate B. The 220 L/min target projects to 26.292–28.074 kPa for A, beyond the gauge. B's 8.4017–9.2779 kPa projection fits both the entered calibration and gauge ranges.
Frequently Asked Questions
Which discharge coefficient belongs here?
Enter a calibrated dimensionless Cd for the same bore, pipe, pressure taps, fluid and installed flow regime. This equation includes the velocity-of-approach factor separately.
Why does the reading show a nominal flow without a flow interval?
The center of the entered pressure reading can lie within the usable range while one uncertainty endpoint falls outside. The worksheet withholds the complete flow interval in that case.
Why can a target be outside the usable range with a numerical result?
The inverse equation gives a conditional pressure projection. The range check compares it with your calibration and gauge. Extrapolated pressures require valid calibration data before use.
Is differential pressure the same as permanent pressure loss?
The worksheet uses the pressure difference between the taps used in calibration. Permanent loss accounts for downstream pressure recovery and depends on other conditions.
Does the calculation give an ISO 5167-compliant installation?
Use the standard and installation documentation for that assessment. This worksheet applies an entered calibration and compares pressure ranges.
Can I use this to meter gas or steam?
The model assumes a steady, incompressible, single-phase liquid. Compressible flow and phase changes require a different model.
Similar Calculators
Volumetric Flow Rate & Collection Audit Calculator
Calculate liquid flow from timed volume or mass collections. Compare measured bounds, duration-weighted flow, meter readings and target fill time.
Valve Cv Calculator & Catalog Travel Window
Compare entered valve Cv curves under clean-liquid assumptions. See where catalog travel may meet a flow target at an entered pressure drop range.
Pipe Pressure Drop Calculator
Compare pipe pressure loss across inside diameters. Calculate Darcy friction, fitting losses, velocity and Reynolds number with explicit transition limits.
Measured Pump Head Calculator & Tap Audit
Calculate pump head from recorded suction and discharge pressures, pipe sizes, elevations and flow. Compare measured head and entered loss corrections.