Four-Wire Resistivity & Conductivity Calculator
Use signed voltage readings and the forward and reverse current magnitudes from two completed four-wire tests of a specimen with a uniform cross section. Each run fits a resistance slope and offset. Compare the voltage error bands after the entered temperature correction, then see whether the common interval reaches your target.
Compare four-wire resistance measurements
Separate a fitted voltage offset from resistivity and compare two measured error bands at the same reference temperature.
All rows share one nΩ·m scale. A single point shows an interval with equal endpoints.
- Run A16.5–17.5 nΩ·m
- Run B17.5–18.5 nΩ·m
- Entered target17–17.5 nΩ·m
- Shared inside target17.5 nΩ·m
Review measured values and exact CSV
| Run | Resistance mΩ | Offset µV | Reference nΩ·m | Conductivity MS/m range |
|---|---|---|---|---|
| Run A | 17/10 | 30 | 33/2 to 35/2 | 400/7 to 2000/33 |
| Run B | 9/5 | 17 | 35/2 to 37/2 | 2000/37 to 400/7 |
How to use this calculator
- Enter the measured sense length and specimen cross section. Enter the documented temperature coefficient and a shared reference temperature.
- Enter both currents as positive magnitudes. Retain the signed voltages from each instrument run, and enter the absolute error per voltage reading.
- Enter the two target endpoints. Compare the resulting signed resistivities and inspect the current and voltage plot for the fitted offsets.
- Review the exact interval or worksheet, including the inverse conductivity where it is defined.
The Formula
The mathematical logic powering this calculation is:
The forward and reverse current magnitudes are positive I₊ and I₋. The corresponding voltages V₊ and V₋ retain their instrument signs. Under an affine voltage model, R = (V₊ − V₋)/(I₊ + I₋) and the offset at zero current is V₊ − I₊R. Microvolts divided by milliamps give milliohms.
For a uniform cross section A and sense length L in millimeters, bulk resistivity in nΩ·m is 1,000 × R(mΩ) × A(mm²) / L(mm). The entered linear model says ρ(T) = ρ(ref) × [1 + (α/100)(T − ref)]. Divide by this factor to normalize each run to the common reference. The factor must be strictly positive.
If each signed voltage is independently within the entered symmetric error e, the smallest slope comes from V₊ − e and V₋ + e. The largest uses V₊ + e and V₋ − e. Equivalently the resistance interval has radius 2e/(I₊ + I₋). Multiply the endpoints by the same positive geometry and temperature correction. Other readings and dimensions are held fixed. Compare both closed intervals and the positive target band. An exact shared endpoint counts as a fit.
Conductivity in MS/m is 1,000 divided by positive resistivity in nΩ·m. The endpoints reverse when taking the reciprocal. A nominal resistivity at or below zero has no conductivity value. A band touching or crossing zero has no finite complete conductivity band under this model. Preserve the signed observations for review.
Length and cross section are 0.001–10,000 mm and mm² respectively. Current magnitudes are 0.001–1,000 mA. Signed voltages are −1,000,000 to 1,000,000 µV and error per voltage is 0–100,000 µV. Temperatures are −50 to 120 °C and the entered coefficient is −5 to 5 percent per °C. Target resistivity is 0.001–10,000,000 nΩ·m. Inputs accept up to six decimal places. Calculated bounds use exact fractions.
Use a uniform bulk sample of known dimensions. The specified error covers entered voltages only. Current calibration, material changes, thermal drift across a polarity pair, temperature-dependent geometry, and sheet resistance require separate treatment.Worked Example
If you input:
- Specimen: 1,000 mm sense length and 10 mm² cross section
- Reference: 20 °C, with both runs measured at 20 °C
- Run A: 100 mA each direction, 200 and −140 µV, 5 µV error
- Run B: 110 and 90 mA, 215 and −145 µV, 5 µV error
Result:
Run A bounds 16.5–17.5 nΩ·m. Run B bounds 17.5–18.5 nΩ·m. Both meet the entered target at exactly 17.5 nΩ·m. The offsets are 30 and 17 µV.
Frequently Asked Questions
Why does the reverse current field use a positive magnitude?
The formula places its observation at a negative current. The measured reverse voltage still needs its displayed plus or minus sign.
Why do the two voltages not need the same magnitude?
They can have different current magnitudes and a constant offset. The model estimates the slope from both points and reports their zero-current intercept.
Can I combine the ends of the two bands as a confidence interval?
The intersection describes resistivity values compatible with both entered voltage-error ranges. It does not assign statistical confidence or model uncertainty in other measurements.
Can I use a film's sheet resistance?
This model expects a bulk specimen of known uniform area and sense length. Sheet resistance and finite probe geometries use different relationships.
Similar Calculators
Ohm's Law Calculator
Calculate voltage, current, or resistance using Ohm's Law — enter any two values to find the third.
Equivalent Resistance Calculator
Solve a DC resistor network, including series, parallel and bridge connections. Inspect node voltages, branch currents, power and terminal tolerance bounds.
DC Voltage Drop & Cable Loss Calculator
Compare DC cable voltage drop, power losses and load limits with separate conductor lengths, connection resistance, constant-power loads and visual sweeps.
Thermal Contact Resistance & Conductivity Fit Calculator
Fit steady thermal resistance versus measured thickness. Estimate apparent conductivity and combined contact resistance, then check held-out runs.
PCB Trace Resistance & Voltage Drop Calculator
Compare PCB copper trace resistance and DC voltage drop across two width layouts, including neckdowns and documented fabrication allowances.