Thermal Contact Resistance & Conductivity Fit Calculator

Use steady measurements at several thicknesses of the same material and test conditions. Separate the resistance that scales with thickness from the combined face-contact resistance. Compare a through-zero fit and review runs withheld from the fit.

Steady heat conduction

Separate conductivity from contact resistance

Edit area and measured runs

Use the same material and metered area with comparable mean temperature, surface conditions and contact loading. Enter bypass only when that known loss is included in metered power.

Run 1
Run 2
Run 3
Run 4

Apparent conductivity

2 W/(m·K)

Combined contact fit

0.001 m²·K/W

Conductivity estimate available

Fitted slope 0.5 m·K/W

Through-zero conductivity 1.4 W/(m·K)

Fit RMSE 0 m²·K/W

Resistance versus thickness

Measured resistance and fitted lines by thicknessThe fitted line can have a nonzero intercept. The amber through-zero line illustrates the result of ignoring contact. Circles are fit runs. Diamonds are check-only runs.Resistance (10⁻³ m²·K/W)03.557.104.529.04Thickness (mm)
Fit runCheck runThrough-zero fit

The fitted line is dashed beyond the thicknesses used to fit it.

Run audit

Resistance columns use 10⁻³ m²·K/W. Residual is measured minus fitted.

RunActualFittedResidual
2 mmFit220
4 mmFit330
6 mmFit440
8 mmCheck · extrapolated5.250.2
Conductivity by run

These separate estimates assign the full resistance of each run to its specimen.

  • Run 1 1 W/(m·K)
  • Run 2 1.33333 W/(m·K)
  • Run 3 1.5 W/(m·K)
  • Run 4 1.53846 W/(m·K)
Select fit CSV

How to use this calculator

  1. Use steady measurements of the same material with a common metered area and comparable mean temperature, contact loading and surface preparation.
  2. For each run, enter thickness, positive hot-to-cold temperature drop, metered watts and any independently determined bypass watts included in that power.
  3. Keep at least two fit runs at distinct thicknesses. Mark other runs Check only to see their residual without altering the fit.
  4. Compare the fitted apparent conductivity with the through-zero estimate. Review the signed slope, combined surface resistance and residuals.
  5. Use the CSV to inspect measured, predicted and exact residual quantities. Check-only points outside the training thickness range are marked as extrapolations.

The Formula

The mathematical logic powering this calculation is:

For common metered area A in square metres and thickness L in metres, use independently corrected conducted power Q = P_metered − P_bypass > 0. Use the positive temperature drop ΔT across the tested specimen-and-contact stack. The areal thermal resistance is R'' = AΔT/Q. Temperature differences in kelvin and degrees Celsius have the same numerical size. A single-run estimate L/R'' absorbs any unseparated contact effects.

Fit R'' = b + mL by ordinary least squares over rows marked Fit. The slope m is centered covariance divided by centered thickness variation. The intercept b is mean resistance minus m times mean thickness. For m>0 and b≥0, apparent conductivity is 1/m and b estimates the combined areal resistance of both interfaces. The fit excludes Check only rows, then predicts both sets from their entered thicknesses. A residual is measured R'' minus predicted R''. RMSE uses the mean of squared residuals over the fit rows. R² uses the centered resistance variation and is unavailable if that variation is zero.

The through-zero comparison uses slope Σ(LR'')/Σ(L²) on the same fit rows, then takes its reciprocal. It shows the result of assuming the contact term is zero. Fewer than two fit rows or identical fit thicknesses leave the free-intercept fit unavailable. A nonpositive slope or negative intercept is reported separately and withholds a physical conductivity estimate. The mathematical line and residuals remain available for diagnosis. Exact fractions carry conversions, slopes, contacts and residuals; the chart and square-root RMSE use rounded coordinates.

Inputs allow two to eight runs, common metered area greater than zero through 1,000,000 cm², positive thickness through 10,000 mm, positive temperature drop through 1,000 K, and positive metered power through 1,000,000 W. Bypass allows zero and must remain below metered power. Decimal fields allow six places. Use one-dimensional steady flow through the same material under comparable mean temperature, pressure, contact and specimen preparation conditions. A nonzero intercept may include fixture effects. The model does not estimate bypass, test whether steady state was achieved, report a calibration uncertainty, or determine compliance with ASTM D5470.

Worked Example

If you input:

  • Metered area: 100 cm²
  • Fit runs: 2, 4 and 6 mm with drops of 1, 1.5 and 2 K
  • Check run: 8 mm and 2.6 K
  • Power per run: 5.25 W metered with 0.25 W bypass

Result:

The exact slope is 0.5 m·K/W and the combined contact resistance is 0.001 m²·K/W, giving 2 W/(m·K). A fit forced through zero gives 1.4 W/(m·K). The 8 mm check run is 0.0002 m²·K/W above the extrapolated fit.

Frequently Asked Questions

Why measure more than one thickness?

Resistance from the specimen scales with thickness, while the model treats combined contact resistance as constant. A thickness series separates their contributions.

What should I enter for temperature drop?

Use the positive steady drop across the stated measurement boundaries, including the two contacts you intend the intercept to capture. Keep those boundaries consistent across runs.

Where does bypass power come from?

Enter measured or otherwise established bypass power already included in metered watts. If your input is net through-specimen heat flow, enter zero bypass.

What does a nonpositive slope mean?

The measured resistance stayed flat or fell with thickness in the unconstrained fit. This model withholds a physical conductivity estimate because a positive slope is required.

What does a negative intercept mean?

The fitted zero-thickness resistance is below zero. Review the measurement conditions, shared interface assumption and power corrections before interpreting the slope and intercept physically.

Why can a check run be an extrapolation?

Its thickness may fall outside the fitted range. It remains excluded from the fit and is compared against a projection beyond the measured training thicknesses.