Vital Lung Capacity & Repeatability Calculator

Enter low and high volume readings from recorded slow vital capacity trials. Mark the trials the operator included, compare the largest two, and calculate volume components for trials with a documented stable end-expiratory baseline.

Recorded slow spirometry

Review recorded vital capacity trials

The two largest included capacities are within the entered tolerance.
Largest included vital capacity
3.6 L
3 of 4 trials included
Mean inspiratory capacity
2.325 L
2 included trials have a supplied baseline
Recorded trial spans

Each bar starts at its trial’s low point. Bar lengths are comparable across trials.

  • T1 · Included3.6 L

    Below baseline 1.2 L · Above baseline 2.4 L

  • T2 · Included3.45 L

    Below baseline 1.2 L · Above baseline 2.25 L

  • T3 · Included3.4 L

    No stable baseline was supplied.

  • T4 · Excluded4.2 L

    Below baseline 1.2 L · Above baseline 3 L

Entered tolerances and volumes

Enter trial code, low and high volumes in liters, an optional baseline, and Y or N for inclusion. Use commas or tabs. Leave a blank column when the baseline is unavailable.

Review recorded trials and CSV
Recorded volume readings and derived intervals
TrialIncludedLowHighBaselineVital capacityICERV
T1Yes0.1 L3.7 L1.3 L3.6 L2.4 L1.2 L
T2Yes-0.2 L3.25 L1 L3.45 L2.25 L1.2 L
T3Yes0 L3.4 LUnavailable3.4 LUnavailableUnavailable
T4No0 L4.2 L1.2 L4.2 L3 L1.2 L

How to use this calculator

  1. Enter the absolute and relative tolerances recorded for your comparison.
  2. Paste each trial code, low and high readings in liters, optional stable end-expiratory baseline, and Y or N for operator inclusion.
  3. Inspect the trial spans and the separation of the two largest included results. Open the ledger for IC and ERV where a baseline was supplied.
  4. Download the exact trial and tolerance worksheet.

The Formula

The mathematical logic powering this calculation is:

For each recorded slow vital capacity trial, subtract the lower volume reading from the higher reading on the same increasing volume scale. This difference is vital capacity (VC) in liters. With a measured stable end-expiratory baseline between the two readings, inspiratory capacity (IC) is high minus baseline and expiratory reserve volume (ERV) is baseline minus low. IC + ERV equals VC. An arbitrary offset applied to all three readings leaves these differences unchanged.

Use only trials marked Y when ranking by VC. The two largest included values determine the gap. The worksheet compares the gap with the smaller of the entered absolute tolerance in liters and the entered relative tolerance multiplied by the largest included VC. Equality is within tolerance. At least three included trials are needed for the comparison. Missing baselines leave the corresponding IC and ERV blank. The mean IC uses included trials with a supplied baseline.

The operator supplies the inclusion flags and any valid stable baseline. The 2019 ATS/ERS spirometry statement explains slow maneuvers, technical acceptability and repeatability. The worksheet checks the entered numerical tolerance only. Test technique, interpretive limits, reference predictions, and clinical decisions require the original pulmonary function report and its clinical interpretation.

Use one to eight trial codes of up to twelve characters beginning with a letter. Volume readings are signed coordinates from −25 to 25 L with ordered endpoints; the baseline may be blank or inside the endpoints. Absolute tolerance is 0–25 L and relative tolerance 0–100%. Input precision is up to six decimals and calculations are exact. The example readings illustrate arithmetic. The CSV preserves the exact values and inclusion state.

Worked Example

If you input:

  • Included spans: 3.600, 3.450 and 3.400 L
  • Excluded trial: 4.200 L
  • Entered tolerance: 0.150 L absolute and 10% relative

Result:

The two largest included spans differ by exactly 0.150 L. The tighter tolerance is 0.150 L. When the first span is 3.601 L, the gap is 0.151 L.

Frequently Asked Questions

How is vital capacity different from total lung capacity?

Vital capacity is the difference between maximum and minimum volume during a recorded maneuver. Absolute total lung capacity includes residual gas and requires an appropriate measurement method.

What does the baseline change?

It splits a trial's measured span into IC and ERV. The total span depends only on the two endpoint readings.

Why is a large excluded trial still visible?

Its recorded value remains available for audit. The included group alone determines the maximum, tolerance check and reported IC average.

What does the repeatability comparison describe?

It checks the numerical gap between included trials against the entered threshold. Clinical interpretation uses the complete study and reference context.