Uncertainty Guard Band & Conformity Calculator
Review up to 20 readings with separate standard uncertainties in one common measurement unit. Compare a center-only check with a three-state interval check, see exactly which readings touch or cross either tolerance limit, and inspect the largest multiplier that keeps each interval contained.
See what uncertainty changes at the tolerance limits
Compare each measured value with its uncertainty interval under the same limits.
2 of 7 readings change classification when the interval is used.
The shaded range covers 9.5 to 10.5. Each marker is the measured value.
- Within limits
- Touches or crosses
- Outside limits
Review guarded limits and CSV
| Reading | Measured value | Center check | Measured interval | Interval check | Guarded acceptance band | Largest contained multiplier |
|---|---|---|---|---|---|---|
| 1 | 9.7 | Within limits | 9.5 to 9.9 | Within limits | 9.7 to 10.3 | 2 |
| 2 | 10 | Within limits | 9.76 to 10.24 | Within limits | 9.74 to 10.26 | 25/6 |
| 3 | 10.3 | Within limits | 10.1 to 10.5 | Within limits | 9.7 to 10.3 | 2 |
| 4 | 10.31 | Within limits | 10.11 to 10.51 | Touches or crosses a limit | 9.7 to 10.3 | 19/10 |
| 5 | 10.6 | Outside limits | 10.5 to 10.7 | Touches or crosses a limit | 9.6 to 10.4 | Center outside |
| 6 | 10.701 | Outside limits | 10.501 to 10.901 | Outside limits | 9.7 to 10.3 | Center outside |
| 7 | 9.5 | Within limits | 9.5 to 9.5 | Within limits | 9.5 to 10.5 | No finite limit |
How to use this calculator
- Enter the lower and upper tolerance limits and the multiplier for the desired interval.
- Paste readings and their standard uncertainty values as two comma- or tab-separated columns in the same unit.
- Compare center-only classification with interval classification. Switch to changed readings to inspect cases affected by uncertainty.
- Open the ledger for each exact guarded acceptance band and its limiting multiplier. Download the complete worksheet as CSV.
The Formula
The mathematical logic powering this calculation is:
With entered lower and upper tolerance limits L and H, a center-only check includes measured value x when L ≤ x ≤ H.
For each entered standard uncertainty u and common multiplier m, the half-width is w = mu. The measured interval is [x − w, x + w]. It is fully within limits if both endpoints lie on or within [L, H]. It lies outside when its upper endpoint is strictly below L or its lower endpoint is strictly above H. Every other case touches or crosses a limit. An interval otherwise outside that touches one limit is counted in the intermediate group. The center-only and interval results use the same closed tolerance limits.
The guarded acceptance band for that row is [L + w, H − w]. A center in this band is equivalent to the whole measured interval staying within tolerance. The acceptance band is empty if the upper bound falls below the lower bound, and it contains one point if they are equal.
For a center within tolerance and u greater than zero, the maximum multiplier that still keeps its interval inside is min(x − L, H − x)/u, inclusive. A zero-uncertainty center within tolerance has no finite cutoff. A center outside tolerance has no feasible multiplier. The calculated cutoff may exceed the available input range.
The worksheet compares a stated interval-based convention with a center-only convention. The chosen multiplier does not determine coverage probability by itself. Apply the rule and uncertainty method documented for the measurement, and consider unresolved cases separately. The calculator supplies no decision about lot acceptance, accreditation or compliance with a specific standard.
Enter one through twenty measured rows. Limits and centers accept ±1,000,000 and standard uncertainty accepts 0–1,000,000, all in a consistent unit. The multiplier accepts 0–10. Values accept up to six decimal places; comparisons are exact rational arithmetic. The CSV contains exact fractions.Worked Example
If you input:
- Tolerance limits: 9.5 to 10.5 in one common measurement unit
- Interval multiplier: 2
- Readings: 7 measurements, including 9.7 ± 0.1 standard uncertainty and 10.6 ± 0.05 standard uncertainty
Result:
Four intervals are fully within, two touch or cross a limit, and one lies outside. At factor 2.01, two are fully within and five touch or cross.
Frequently Asked Questions
Can two readings have different guard bands?
Yes. The common multiplier is multiplied by each row's own standard uncertainty. Larger standard uncertainty contracts its guarded acceptance band more.
Does touching a tolerance limit count as within?
If the whole interval remains within the tolerance, equality at either bound is accepted. If the rest of an interval lies outside and touches one bound, it goes in the intermediate group.
What if the uncertainty interval is wider than the tolerance?
The row cannot be fully contained at that multiplier. Its guarded acceptance band is empty. The interval can still overlap the tolerance or lie wholly outside it.
Can I enter expanded uncertainty with its factor already applied?
Yes, by using a multiplier of one and pasting that interval half-width in the uncertainty column. Record the underlying expanded uncertainty method with the exported report.
Similar Calculators
Gauge Accuracy & Linearity Calculator
Audit gauge accuracy with five reference values and three repeat readings per value. Compare signed bias, repeatability, linearity and lack of fit.
Uncertainty Propagation Calculator
Propagate independent measurement uncertainties through sums, products and ratios. Compare contributions, expanded uncertainty and input improvement scenarios.
Process Capability & Cpk Calculator
Compare Cp, Cpk and Cpm for two processes. Enter sample readings or summaries, set specification limits and see which change would meet your index goal.
Dimensional Tolerance Stack-Up & Fit Calculator
Add or subtract up to six dimensional tolerances. Compare worst-case clearance with an exact modeled fit probability, signed contributions and the full gap chart.
Vital Lung Capacity & Repeatability Calculator
Calculate vital capacity from recorded slow spirometry endpoints. Compare included trials, entered repeatability tolerances and optional inspiratory capacity.