Raman Shift & Detector Window Calculator

Enter one target Raman shift magnitude, its allowed spread and two excitation and detector ranges. Compare the resulting signed shift windows, inspect exact overlap on each branch and find the largest detector-only half-width around each nominal line.

Raman spectroscopy · Detector coverage

Compare Raman shift detector windows

Edit the target and detector setups

Use one documented continuous usable band per setup. Enter all wavelengths on the same vacuum basis.

Setup A
Setup B

Only setup A covers both target bands.

Target magnitude 3,750 ± 250 cm⁻¹

Setup A · 500 nm excitation

−5,000 to 4,000 cm⁻¹

Usable signed shift window

Stokes
Full target band
Anti-Stokes
Full target band

Setup B · 625 nm excitation

−4,000 to 3,500 cm⁻¹

Usable signed shift window

Stokes
Touches one boundary
Anti-Stokes
Full target band

Inspect a Raman branch

Detector windows and target bands

Signed Raman shift · cm⁻¹

  • Usable window
  • Target band
  • Shared shift on selected branch

Stokes · 3,750 cm⁻¹

Target band
3,500 to 4,000 cm⁻¹

Setup A · Full target band

Shared target shift
3,500 to 4,000 cm⁻¹
Maximum detector half-width
250 cm⁻¹
Nominal scattered wavelength
615.384615 nm
Target wavelength endpoints
606.060606 nm to 625 nm

Setup B · Touches one boundary

Shared target shift
3,500 cm⁻¹
Maximum detector half-width
Center outside window
Nominal scattered wavelength
816.326531 nm
Target wavelength endpoints
800 nm to 833.333333 nm
Review the exact spectral ledger
Detector branch limits and conversions
Setup and branchTarget shiftUsable shift windowCoverageShared target shiftNominal wavelength
A · Stokes3,500 to 4,000 cm⁻¹−5,000 to 4,000 cm⁻¹Full target band3,500 to 4,000 cm⁻¹615.384615 nm
A · Anti-Stokes−4,000 to −3,500 cm⁻¹−5,000 to 4,000 cm⁻¹Full target band−4,000 to −3,500 cm⁻¹421.052632 nm
B · Stokes3,500 to 4,000 cm⁻¹−4,000 to 3,500 cm⁻¹Touches one boundary3,500 cm⁻¹816.326531 nm
B · Anti-Stokes−4,000 to −3,500 cm⁻¹−4,000 to 3,500 cm⁻¹Full target band−4,000 to −3,500 cm⁻¹506.329114 nm

How to use this calculator

  1. Enter the nominal target magnitude and the permitted shift half-width.
  2. Enter each excitation wavelength and its usable continuous detector range on the same wavelength basis.
  3. Read the shared band and remaining tolerance for Stokes or anti-Stokes. Export the exact overlap and converted wavelengths.

The Formula

The mathematical logic powering this calculation is:

Convert every vacuum wavelength λ in nm to a wavenumber of 10⁷/λ in cm⁻¹. The signed Raman shift is the excitation wavenumber minus the scattered-light wavenumber. Stokes is positive with longer scattered wavelength. Anti-Stokes is negative with shorter scattered wavelength. Use a consistent vacuum wavelength basis across all inputs.

For each setup, the lower and upper usable wavelengths map to the lower and upper signed shifts. Compare the documented window with two target bands. Stokes is centered at the entered magnitude and anti-Stokes at its negative. Subtract and add the allowed half-width to each center. Exact contact is inside a closed interval. A shared endpoint alone is one-point contact when the target has positive width. A zero-width target inside the window qualifies in full.

The maximum detector-only half-width is the smaller distance from the signed center to the two detector shift bounds. It is unavailable when the center is outside. This value measures geometric room on both sides of the center. It may exceed the entered shift magnitude. The entered allowance is limited to the magnitude so each target remains on its own side of zero or touches zero.

To translate target endpoints into wavelengths, subtract each signed shift from the excitation wavenumber and invert the result. A scattered wavenumber at or below zero has no positive finite wavelength, so that endpoint displays as unavailable. The coverage check still uses the detector’s finite shift interval.

Inputs use up to three decimals. Excitation and usable detector wavelengths accept 100–10,000 nm, and equal detector endpoints describe one wavelength. The target magnitude is 0.001–100,000 cm⁻¹. Exact rational comparisons establish coverage, while decimal results and the plot round for reading. CSV cells preserve integers or exact fractions. Enter one documented continuous usable band per setup that already reflects the relevant optics and filters. This worksheet compares geometric coverage. Intensity, resolution, calibration uncertainty and instrument operation are separate.

Worked Example

If you input:

  • Target magnitude and allowance: 3,750 ± 250 cm⁻¹
  • Setup A: 500 nm excitation, usable detector 400–625 nm
  • Setup B: 625 nm excitation, usable detector 500–800 nm

Result:

A spans −5,000 to +4,000 cm⁻¹ and covers both branches. B spans −4,000 to +3,500 cm⁻¹. Its anti-Stokes band fits exactly at the lower end and its Stokes band only touches at +3,500 cm⁻¹.

Frequently Asked Questions

Why does Stokes have a positive shift but a longer wavelength?

Stokes scattering has a lower wavenumber than the excitation. Subtracting it from the excitation produces a positive shift. Lower wavenumber corresponds to longer wavelength.

Do these limits guarantee that a peak can be resolved?

The worksheet checks whether its wavelength is inside the entered usable band. Resolving peaks also depends on spectral resolution, signal and calibration.

Can I check a single wavelength or a single shift?

Yes. Equal detector endpoints define one wavelength. Zero allowed half-width defines one signed shift on each branch.

Why is maximum half-width sometimes unavailable?

The signed nominal center lies outside that setup’s usable range. An allowance around that center cannot produce full coverage there.

How do I check a measured wavelength interval?

Enter its endpoints as the usable lower and upper wavelengths. The setup result returns the corresponding exact signed Raman shift interval.