Numerical Aperture & Microscope Camera Sampling Calculator
Use a stated objective NA or calculate it from the refractive index and acceptance half-angle. Compare two camera configurations with independent pixel pitches, sensor sizes, binning and relay magnifications. See which camera Nyquist frequencies cover the ideal optical bandwidth and how much of the specimen each setup captures.
Compare microscope optics and camera sampling
Model numerical aperture, specimen coverage and the camera sampling limit for two optical relay setups.
Camera A covers the optical cutoff.
NA 0.8 · optical cutoff 2.909091 cycles/µm.
Rayleigh lateral separation 0.419375 µm · sample spacing limit 0.171875 µm/sample.
Camera A
0.1625 µm/sample
40× at sensor · 3.076923 cycles/µm at Nyquist.
Nyquist exceeds the optical cutoff
Projected field 332.8 × 249.6 µm.
Output 2,048 × 1,536 samples.
Minimum relay for this camera’s pitch and binning 0.945455×.
Camera B
0.325 µm/sample
40× at sensor · 1.538462 cycles/µm at Nyquist.
Nyquist falls below the optical cutoff
Projected field 332.8 × 249.6 µm.
Output 1,024 × 768 samples.
Minimum relay for this camera’s pitch and binning 1.890909×.
Incomplete bin edges omit 1 native column and 1 native row.
Optical contrast and camera Nyquist
1.454545 cycles/µm · 39.100222% ideal optical contrast.
- Camera A covers this frequency.
- Camera B covers this frequency.
Projected specimen field
Edit microscope optics
Edit camera setups
Use square pixels, the active native sensor dimensions and optical relay magnification.
Review exact sampling calculations
Optical cutoff 32/11 cycles/µm.
Camera A
Sensor magnification 40 × 1 = 40×.
Sample spacing 1 × 13/2 ÷ 40 = 13/80 µm/sample.
Nyquist 40/13 cycles/µm.
Exact projected field 1664/5 × 1248/5 µm.
Camera B
Sensor magnification 40 × 1 = 40×.
Sample spacing 2 × 13/2 ÷ 40 = 13/40 µm/sample.
Nyquist 20/13 cycles/µm.
Exact projected field 1664/5 × 1248/5 µm.
Show sampling CSV text
How to use this calculator
- Enter the detected wavelength and stated objective magnification.
- Select stated NA or enter the medium’s refractive index and half-angle.
- Record native sensor pitch and dimensions, optical relay magnification and binning for both camera setups.
- Compare digital Nyquist with the optical cutoff and inspect the field projected onto each sensor.
- Download settings and exact sampling calculations.
The Formula
The mathematical logic powering this calculation is:
Enter the objective's stated numerical aperture (NA), or derive it as n sin α from the refractive index n and acceptance half-angle α. Use the detected wavelength λ in air or vacuum. Convert nanometers to micrometers before comparing optical and digital spatial frequencies.
For an ideal diffraction-limited incoherent optical system with a circular pupil, the cutoff frequency is f_c = 2NA/λ. The Rayleigh lateral point separation is 0.61λ/NA. Optical modulation at frequency f = u f_c is (2/π)[arccos(u) − u√(1 − u²)] for 0 ≤ u ≤ 1, and zero beyond that limit. The plotted curve shows ideal optical contrast transfer. Real objectives and specimen contrast affect measured response.
For each camera, sensor magnification M is objective magnification multiplied by that camera's entered optical relay factor. If native sensor pitch is p and the camera groups b native pixels along each axis, the sample-plane spacing is s = bp/M micrometers per sample. Nyquist frequency is 1/(2s). Compare it with f_c, which corresponds to a sample spacing limit of λ/(4NA). The minimum relay to reach that limit is bp divided by the product of the spacing limit and the objective magnification. A digital zoom or resampling setting does not change optical projection onto the sensor.
For native columns C and rows R, the complete output groups are floor(C/b) across and floor(R/b) high. Multiply these counts by s for the rectangular specimen field. The model records any omitted edge pixels. Pixel pitch, magnification, projected spacing, Nyquist and the rectangular field use exact decimal arithmetic. Stated NA gives an exact cutoff comparison. With angle-derived NA, the classification uses a floating-point roundoff allowance.
The model covers monochromatic incoherent imaging, square pixels and uniform magnification across the rectangular field. Partial coherence, optical aberrations, illumination, relay image-circle coverage, vignetting, detector contrast, color filters and noise are separate considerations.Worked Example
If you input:
- Optics: 550 nm, 40× objective, NA 0.8
- Camera A: 6.5 µm native pitch, 1× relay, no binning
- Camera B: 6.5 µm native pitch, 1× relay, 2×2 binning
Result:
The ideal optical cutoff is about 2.909 cycles/µm. Camera A projects 0.1625 µm per sample and covers the cutoff. Camera B projects 0.325 µm per sample and falls below it. The two retained specimen fields are both 332.8 × 249.6 µm.
Frequently Asked Questions
Is numerical aperture the same as objective magnification?
Numerical aperture is the product of refractive index and sine of the acceptance half-angle. Optical magnification determines how the sample projects onto a sensor. These inputs affect different parts of the calculation.
How does binning change the captured field?
Binning makes each sample span more specimen area and reduces the output pixel count. The field stays the same for evenly divisible native dimensions with the same relay. The model omits incomplete edge bins when they occur.
Why can two cameras capture the same area with different sampling?
Their output pixel counts may differ while their projected fields match. A coarser sample spacing lowers the Nyquist frequency even when the same amount of specimen remains visible.
Does covering the cutoff guarantee sharp images?
The comparison matches ideal diffraction bandwidth to digital sampling. Contrast loss and optical effects such as aberration, illumination and vignetting are still relevant.
What is camera relay magnification?
Use the optical factor between the objective system and the sensor. Digital cropping, preview scaling and monitor magnification do not change sample-plane pixel spacing.
Similar Calculators
Camera Field of View & Framing Calculator
Compare camera field of view across object distances, focal lengths and sensor crops. Check width and height limits and solve the focal length range that fits.
Optical Lens Thickness & Sag Planner
Calculate edge thickness and both surface sags for two spherical optical lens geometries. Compare cross sections and test an entered minimum thickness.
Wavelength & Phase Path Calculator
Calculate frequency and wavelength in up to four media. Compare path delay, wrapped phase and the shared frequency range for entered lengths.
Photo Print Size & Crop Calculator
Compare print sizes, native pixels per inch, and fit or fill cropping without resampling. Preview an optional local photo and calculate the minimum dimensions for a printer's target PPI.
